Volume 20, Issue 3 (Pajouhan Scientific Journal, Summer 2022)                   Pajouhan Sci J 2022, 20(3): 127-132 | Back to browse issues page

Research code: 9906184036
Ethics code: IR.UMSHA.REC.1397.879


XML Persian Abstract Print


Download citation:
BibTeX | RIS | EndNote | Medlars | ProCite | Reference Manager | RefWorks
Send citation to:

Babaei M, Khalili M, Khalili A, Sharifian P. Evaluation of Suitable Connection Point of Skin Probe for Temperature Control in Premature Infants under Radiant Heating Admitted to Neonatal Intensive Care Unit. Pajouhan Sci J 2022; 20 (3) :127-132
URL: http://psj.umsha.ac.ir/article-1-807-en.html
1- Department of Nursing, Tuyserkan Branch, Islamic Azad University, Tuyserkan, Iran
2- Department of Pediatric Nursing, Yasouj University of Medical Sciences, Yasouj, Iran
3- Mother and Child Care Research Center, Hamadan University of Medical Sciences, Hamadan, Iran
4- Nursing Care Research Center (NCRC), School of Nursing and Midwifery, Iran University of Medical Sciences, Tehran, Iran , Rsharifian7@gmail.com
Abstract:   (1133 Views)
Background and Objectives: Placement of skin heat probe for continuous monitoring of infant body temperature is one of the main components of neonatal care. Therefore, managing the body temperature of newborns under radiation warmers requires a lot of attention from nurses to prevent unwanted complications. The aim of this study was to determine the appropriate connection point of skin probe for temperature control in premature infants under radiant heating admitted to neonatal intensive care units in Hamadan in 2020.
Materials and Methods: In this cross-sectional descriptive study, 140 infants aged 34 to 37 weeks admitted to neonatal intensive care units in Hamadan in 2020 who met the inclusion criteria were included in the study by available methods. The incubator temperature was set at 37 degrees and it was the same for all infants. For each infant, the probe was attached to the forehead, chest between the ribs, right axillar, arms adjacent to the nipple, right armpit, back of the legs, and upper thigh, respectively. The baby's temperature displayed on the digital warmer screen was recorded two minutes later. A researcher-made checklist was used to record the data and the data were analyzed using t-test and Paired t-test.
Results: The correlation between left and right axillary was highest, meaning that the best location for the incubator thermometer probe, after the left axillary is the right axillary (r = 0.91, P = 0.001).
Conclusions: The most suitable place for connecting skin probe for temperature control after left axillary, is the right axillary.
Full-Text [PDF 462 kb]   (733 Downloads)    
Type of Study: Research Article | Subject: Basic Sciences
Received: 2021/09/19 | Accepted: 2021/10/9 | Published: 2022/09/10

References
1. Dubos C, Delanaud S, Brenac W, Yassin FC, Carpentier M, Tourneux P. The newborn infant's thermal environment in the delivery room when skin-to-skin care has to be interrupted. J Matern Fetal Neonatal Med. 2020:1-7. [DOI:10.1080/14767058.2020.1838479] [PMID]
2. George G, Mishra S. Routine axillary temperature monitoring in neonates cared under radiant warmer - is it necessary? Indian J Pediatr. 2009;76(12):1281-2. [DOI:10.1007/s12098-009-0234-7] [PMID]
3. Gopalakrishnan S, Karmani S, Ramar P, Pandey A, Sodhi K. Conductive thermal mattress versus routine care to reduce neonatal hypothermia during transport among low-birthweight neonates: An experimental study with historical controls. MJAFI. 2021; In Press. [DOI:10.1016/j.mjafi.2020.10.022] [PMID] []
4. Koh KH, Yu CW. Comparing the accuracy of skin sensor temperature at two placement sites to axillary temperature in term infants under radiant warmers. J Neonatal Nurs. 2016;22(4):196-203. [DOI:10.1016/j.jnn.2016.01.003]
5. Knobel RB. Thermal stability of the premature infant in neonatal intensive care. Newborn Infant Nurs Rev. 2014;14(2):72-6. [DOI:10.1053/j.nainr.2014.03.002]
6. D'Agati SS. The effects of a sound-modified environment on physiological variables in premature infants in neonatal intensive care; 1995.
7. Costeloe K, Hennessy E, Gibson AT, Marlow N, Wilkinson AR. The EPICure study: outcomes to discharge from hospital for infants born at the threshold of viability. Pediatrics. 2000;106(4):659-71. [DOI:10.1542/peds.106.4.659] [PMID]
8. da Mota Silveira SM, de Mello MJG, de Arruda Vidal S, de Frias PG, Cattaneo A. Hypothermia on admission: a risk factorfor death in newborns referred to the Pernambuco Institute of Mother and Child Health. J Trop Pediatr. 2003;49(2):115-20. [DOI:10.1093/tropej/49.2.115] [PMID]
9. Dey K, Deb UK. Modeling and simulation of heat transfer phenomenon from infant radiant warmer for a newborn baby. OJMSi. 2021;9(2):111. [DOI:10.4236/ojmsi.2021.92007]
10. Silverman WA, Fertig JW, Berger AP. The influence of the thermal environment upon the survival of newly born premature infants. Pediatrics. 1958;22(5):876-86. [DOI:10.1542/peds.22.5.876] [PMID]
11. Knobel R, Holditch‐Davis D. Thermoregulation and heat loss prevention after birth and during neonatal intensive‐care unit stabilization of extremely low‐birthweight infants. J Obstet Gynecol Neonatal Nurs. 2007;36(3):280-7. [DOI:10.1111/j.1552-6909.2007.00149.x] [PMID]
12. Hockenberry MJ, Wilson D. Wong's nursing care of infants and children. 10th ed. Philadelphia, PA: Elsevier; 2015.
13. Merchant RM, Topjian AA, Panchal AR, Cheng A, Aziz K, Berg KM, et al. Part 1: executive summary: 2020 American Heart Association guidelines for cardiopulmonary resuscitation and emergency cardiovascular care. Circulation. 2020;142(16_Suppl_2):S337-57. [DOI:10.1161/CIR.0000000000000918]
14. Knobel RB, Holditch-Davis D, Schwartz TA, Wimmer JE. Extremely low birth weight preterm infants lack vasomotor response in relationship to cold body temperatures at birth. J Perinatol. 2009;29(12):814-21. [DOI:10.1038/jp.2009.99] [PMID] []
15. Lyon AJ, Pikaar ME, Badger P, McIntosh N. Temperature control in very low birthweight infants during first five days of life. Arch Dis Child Fetal Neonatal Ed. 1997;76(1):F47-50. [DOI:10.1136/fn.76.1.F47] [PMID] []
16. Thomas KA. Preterm infant thermal responses to caregivingdiffer by incubator control mode. J Perinatol. 2003;23(8):640-5. [DOI:10.1038/sj.jp.7211002] [PMID]
17. Schafer D, Boogaart S, Johnson L, Keezel C, Ruperts L, Vander Laan KJ. Comparison of neonatal skin sensor temperatures with axillary temperature: does skin sensor placement really matter? Adv Neonatal Care. 2014;14(1):52-60. [DOI:10.1097/ANC.0000000000000027] [PMID]
18. Chaseling GK, Molgat-Seon Y, Daboval T, Chou S, Jay O. Body temperature mapping in critica ewborn infants nursed under radiant warmers during intensive care. J Perinatol. 2016;36(7):540-3. [DOI:10.1038/jp.2016.16] [PMID]
19. Trevisanuto D, Coretti I, Doglioni N, Udilano A, Cavallin F, Zanardo V. Effective temperature under radiant infant warmer: Does the device make a difference? Resuscitation. 2011;82(6):720-3. [DOI:10.1016/j.resuscitation.2011.02.019] [PMID]
20. Thomas KA. Comparability of infant abdominal skin and axillary temperatures. Newborn Infant Nurs Rev. 2003;3(4):173-8. [DOI:10.1053/S1527-3369(03)00123-5]
21. Robertson-Smith J, McCaffrey FT, Sayers R, Williams S, Taylor BJ. A comparison of mid-forehead and axillary temperatures in newborn intensive care. J Perinatol. 2015;35(2):120-2. [DOI:10.1038/jp.2014.148] [PMID]
22. Duran R, Vatansever U, Acunaş B, Süt N. Comparisons of temporal artery, mid forehead skin and axillary temperature recordings in preterm infants<1500 g of birthweght. J Paediatr Child Health. 2009; 45(7-8):444-7. [DOI:10.1111/j.1440-1754.2009.01526.x] [PMID]
23. Pouy S, Chehrzad MM. Identification the best skin temperature probe attachment place in premature neonates nursed under radiant warmers in NICU: A diagnostic clinical trial study. J Neonatal Nurs. 2019:25(2):69-73. [DOI:10.1016/j.jnn.2018.10.001]

Add your comments about this article : Your username or Email:
CAPTCHA

Send email to the article author


Rights and permissions
Creative Commons License This work is licensed under a Creative Commons Attribution-NonCommercial 4.0 International License.

© 2024 CC BY-NC 4.0 | Pajouhan Scientific Journal